Code: EML0294
Gel electrophoresis Unit.
AGAROSE GEL ELECTROPHORESIS SYSTEM WITH POWER PACK
For electrophoresis apparatus
· Should have Cell size (W x L x H) 17.8 x 25.5 x 6.8 cm, and 9.2 x 25.5 x 5.6 cm
· Gel tray sizes (OD) (W x L) Should be 15 x 7 cm, 15 x 10 cm, 7 x 7 cm , 7 x 10 cm
· Should have Ready agarose gels accommodated for wide mini and 96 Plus formats, 8-, 12-, 2 x 8-well
· Should have Sample throughput around 10–60
· Should have Base buffer volume approximately ~650 ml
· Should not have Buffer recirculation
· Should have Bromophenol blue migration ~4.5 cm/hr (at 75 V)
For power pack
· Should be suitable for running horizontal and vertical applications.
· Should have control of constant voltage output as well as constant current output.
· Output current should be in the range 10–300 V, fully adjustable in 1 V steps4–400 mA, fully adjustable in 1 mA steps 75 W (maximum) and timer; 1 min–99 hr. 59 min, fully adjustable and crossover Output terminals 4 pair recessed banana jacks in parallel
· Manual programming of voltage, current, temporary changes to power and time parameter can be made in the power pack without interring the run.
· 4 or more frequently used programme should be stored in the system of power pack.
· Should include Safety compliance No-load detection; rapid resistance change detection, ground leak detection, overload/short circuit detection, overvoltage protection, over-temperature protection Input protection Fuse on hot and neutral
· Should be suitable to operate at 220 to 230 V AC, 50 to 60 Hz.
· Should have UV torch with dual Wavelength (365 and 254 nm) for gel viewing and gel rocker for gel staining
· One year comprehensive warranty for the whole system.
· Maintenance back up services in East African region
Supplied with startup consumables