Code: ETOPM10600
LC/MS.
LC specifications:-
Flow rate range = 10 nL/min to 250 nL/min (gradient)
Max. pressure = 20 Mpa 10 nL/min to 1,000 nL/min (conditioning) Retention time reproducibility = RSD 0.5% (Retention time of flow rate 200 nL/min)
Sample injection = (1) Autosampler: standard rack (200 vials) or optional 96-well plate rack; direct injection with
(2) Manual injector (optional)
Dimension/weight = (w/ autosampler) W720 x L470 x H650 mm (85 kg)
(w/o autosampler) W340 x L470 x H650 mm (46 kg)
Power requirement = AC 100~240V ± 10% 50/60Hz Ambient conditions = Humidity: 30~80%; Temperature: 15~35 degree C
Mass Spectrometer Specifications MS mode measurement range = 50 to 25,000 (m/z)
Precursor ion selection range = 150 to 4,000 (m/z)
Resolution = 10,000 (m/z 1,000) or over MS accuracy = 5 ppm with internal reference
Nitrogen gas = Pressure 0.65 MPa; Max. flow rate 2.5 L/min; Purity 99% (with Nitrogen gas generator
Helium gas = Pressure 0.35 MPa; Max flow rate 10 mL/min; Purity 99.9999%
Weight = 925 lb
Dimensions = 47 in (w) x 32 in (d) x 55 in (h)
Power requirement = AC 100~240V ± 10% 50/60Hz
Ambient temperature = 15 to 35 ËšC (-/+ 2 ËšC within a day)
Nitrogen Generator Installation, testing, commissioning and instrument training by qualified engineer during commissioning is required. Supplied with a complete set of operation and maintenance manuals.